Model 240

Operation Manual

OM024-6 Model 240 Operating Manual

Page 1

loading

Related Manuals for AmpliVox 240

Summary of Contents for AmpliVox 240

  • Page 1
    Model 240 Operation Manual OM024-6 Model 240 Operating Manual Page 1…
  • Page 2: About This Manual

    ABOUT THIS MANUAL READ THIS OPERATING MANUAL BEFORE ATTEMPTING TO USE THE INSTRUMENT. This manual is valid for the Model 240 (applies from firmware version 4v47 onwards – please refer to section 1.3). This product is manufactured by: Amplivox Ltd.

  • Page 3
    For supply in US only Caution: Federal Law restricts this device to sale by or on the order of a licenced medical professional.
  • Page 4: Table Of Contents

    Contents 1 Introduction 1.1 Intended applications 1.2 Unpacking 1.3 Firmware version 1.4 Standard contents 1.5 Optional accessories 2 Important Safety Instructions 2.1 Precautions 2.2 Electromagnetic compatibility (EMC) considerations 2.3 Power supply options 2.4 Audiometer connections 2.5 Data transfer to a printer 2.6 Data transfer to a computer 3 Using the Audiometer 3.1 Switching the audiometer on and off…

  • Page 5: Introduction

    1 Introduction Thank you for purchasing an Amplivox audiometer. The Amplivox Model 240 is a diagnostic audiometer that will give many years of reliable service if treated with care. 1.1 Intended applications The Model 240 diagnostic audiometer is designed for use by audiologists, general practitioners, hearing aid dispensers and child health professionals.

  • Page 6: Optional Accessories

    Printer cable(s) USB Cable Audiocups (noise reducing earphone enclosures) 2 Important Safety Instructions The Model 240 instrument must be used only by practitioners qualified to perform audiometric tests. It is intended for use as a screening and diagnostic tool. 2.1 Precautions…

  • Page 7: Electromagnetic Compatibility (Emc) Considerations

    Once the voltage of the batteries is too low to operate the instrument the message “Replace Battery” will appear. Note that local regulations are likely to cover disposal of the batteries. OM024-6 Model 240 Operating Manual Page 3…

  • Page 8: Audiometer Connections

    Patient Response Switch * The relevant part numbers are indicated in Section 12 Note regarding the 6-pin mini DIN connector: This is a restricted socket for Amplivox use only. No user access is permitted. Page 4 OM024-6 Model 240 Operating Manual…

  • Page 9: Data Transfer To A Printer

    The audiometer is supplied with software to allow connection to a computer for the transfer of test results (see Section 3.9). You must use the designated USB cable which is available from Amplivox (see Section 12). OM024-6 Model 240 Operating Manual…

  • Page 10: Using The Audiometer

    This indicates that when the PRESENT key is pressed, a tone will be presented at 30dBHL at a frequency of 1kHz (1000Hz) to the designated ear. On start up the audiometer defaults to the left ear. Page 6 OM024-6 Model 240 Operating Manual…

  • Page 11: Audiometer Controls

    YES key to load the audiogram; then release MENU  Contrast Adjust contrast using the SIGNAL keys Shows either “Not in use” or the battery Battery voltage, e.g. “5.3v”. OM024-6 Model 240 Operating Manual Page 7…

  • Page 12
    Press this key when an automatic test has been completed to displays the results; use the LEFT and RIGHT keys to display the results for the required ear and   the FREQUENCY keys to view the entire frequency range. Page 8 OM024-6 Model 240 Operating Manual…
  • Page 13
    5dB steps; to scroll through the range keep the key pressed PRESENT Press to present the displayed test signal to the patient. The “PRESENT” indicator above the display will illuminate green during presentation OM024-6 Model 240 Operating Manual Page 9…
  • Page 14: Threshold Retention Function

    Left ear 20dBHL Right ear 10dBHL To clear the Threshold Retention memory, use the Clear Test menu option described in Section 3.4.2. To record and review bone conduction thresholds use the BONE key. Page 10 OM024-6 Model 240 Operating Manual…

  • Page 15: Saving Audiograms In Internal Memory

    3.8 Printing audiograms Three designated thermal printers (the Able AP1300 or the Sanibel MPT- II) are available as options for use with the Model 240 audiometer. The correct printer must be selected (use the MENU options described in Section 3.4.2 to make this selection).

  • Page 16: Data Transfer To Noah Or Audiview

    3.9 Data transfer to NOAH or AudiView To transfer test results stored within the audiometer to a NOAH database the Amplivox NOAH Audilink software must be installed on to a computer. Alternatively, Amplivox AudiView allows data to be transferred to a computer and subsequently viewed, annotated &…

  • Page 17
    30dB, increase the attenuation level in 10dB steps until the patient responds (2) When the patient responds, wait for 1 to 2 seconds and present the tone again at the same level; however, if the patient does respond at OM024-6 Model 240 Operating Manual Page 13…
  • Page 18
    Print the results (Section 3.8), or  Transfer the results to a computer (Section 3.9) Refer to Section 3.4.2 to clear the thresholds at the end of a test and, if required, switch off the audiometer. Page 14 OM024-6 Model 240 Operating Manual…
  • Page 19
    (3) To initiate a test press and hold the MENU key, then press the AUTO key and release both OM024-6 Model 240 Operating Manual Page 15…
  • Page 20
    The thresholds are retained by the audiometer and may be viewed, stored, printed or transferred to a computer (see Section 4.4.4). Refer to Section 3.4.2 to clear the thresholds at the end of a test and, if required, switch off the audiometer. Page 16 OM024-6 Model 240 Operating Manual…
  • Page 21
    (e.g. if no response was made and the retry option was not chosen). The operator then has the option to use manual audiometry to obtain any missing thresholds. Press RESULTS to continue. OM024-6 Model 240 Operating Manual Page 17…
  • Page 22: Specification

    USB interface: Transfer of test results to a computer 5.2 Maximum hearing levels provided at each frequency Frequency, Air conduction, dBHL Bone conduction, dBHL (DD45) (B-71) 1000 1500 2000 3000 4000 6000 8000 Page 18 OM024-6 Model 240 Operating Manual…

  • Page 23
    Continuous operation Equipment mobility Portable The Model 240 Audiometer is classified as a Class IIa device under Annex IX of the EU Medical Devices Directive. It is intended for use as a diagnostic audiometer instrument. OM024-6 Model 240 Operating Manual…
  • Page 24: Symbols

    6 to 8 hours use from alkaline batteries. Frequency Modulation Carrier frequencies: 125Hz to 8kHz as per pure tones Modulation waveform: Sinusoidal Rising and falling symmetry: Symmetrical on linear frequency scale Modulating frequency: 15.625Hz Frequency deviation: +/-10% Page 20 OM024-6 Model 240 Operating Manual…

  • Page 25: Routine Maintenance

    8 Routine Maintenance 8.1 Audiometer maintenance The Model 240 audiometer is a precision instrument. Handle it carefully in order to ensure its continued accuracy and service. When cleaning the instrument, first disconnect it from the mains supply. Use a soft cloth and…

  • Page 26
    Before use check the transducer cables and connectors for signs of wear and/or damage. If you find any, please replace the item immediately by contacting Amplivox or your Amplivox distributor, requesting the relevant part number (see Section 12). Handle the audiometric headset, bone vibrator headset and other accessories with care.
  • Page 27
    8.3 Mains adapter maintenance Before use check the mains AC adapter for signs of wear and/or damage. If you find any replace the adapter immediately by contacting Amplivox or your Amplivox distributor. Refer to Section 12 for approved part numbers.
  • Page 28: Instrument Storage And Transportation

    Batteries (if fitted) should be removed if the instrument is not to be used over an extended period of time. 10 Calibration and Repair of the Instrument Amplivox recommends that this audiometer should be calibrated on an annual basis. Please contact Amplivox or the designated distributor for details of calibration services.

  • Page 29: Guarantee

    12 Ordering Consumables and Accessories To order consumables, additional accessories and to replace detachable parts that have been damaged, please contact Amplivox for current prices and delivery charges. The items available are listed below: Stock No.

  • Page 30: Disposal Information

    For any waste electrical units purchased from Amplivox that either:  bear the crossed out wheeled bin symbol with black bar underneath  or, have been replaced with new Amplivox products on a like-for-like basis Page 26 OM024-6 Model 240 Operating Manual…

  • Page 31
    B2B Compliance will be able to provide further information on how to recycle your waste electrical units and answer any queries you may have. B2B Compliance Tel: +44 (0) 1691 676 124 (Option 2) Email: operations@b2bcompliance.org.uk OM024-6 Model 240 Operating Manual Page 27…
  • Page 32: Appendix 1 — Emc Guidance & Manufacturer’s Declaration

    Appendix 1 — EMC Guidance & Manufacturer’s Declaration Guidance and manufacturer’s declaration – electromagnetic emissions The Model 240 Audiometer is intended for use in the electromagnetic environment specified below. The customer or user of Model 240 Audiometer should assure that it is used in such an environment.

  • Page 33
    Guidance and manufacturer’s declaration – electromagnetic immunity (1) The Model 240 Audiometer is intended for use in the electromagnetic environment specified below. The customer or user of the Model 240 Audiometer should assure that it is used in such an environment.
  • Page 34
    IEC 61000-4-8 typical location in a typical commercial or hospital environment. NOTE U is the a.c. mains voltage prior to the application of the test level Page 30 OM024-6 Model 240 Operating Manual…
  • Page 35
    Guidance and manufacturer’s declaration – electromagnetic immunity (2) The Model 240 Audiometer is intended for use in the electromagnetic environment specified below. The customer or user of the Model 240 Audiometer should assure that it is used in such an environment.
  • Page 36
    RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Model 240 Audiometer is used exceeds the applicable RF compliance level above, the Model 240 Audiometer should be observed to verify normal operation.
  • Page 37
    Recommended separation distances between portable and mobile RF communications equipment and the Model 240 Audiometer The Model 240 Audiometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Model 240 Audiometer can help prevent electromagnetic…
  • Page 38: Appendix 2 — Use With Non-Medical Electrical Equipment

    16 of IEC 60601-1:2005 are met. The following signal inputs and outputs on the Model 240 audiometer are electrically isolated to the requirements of IEC 60601-1 in order to reduce any potential hazard associated with the use of mains-powered equipment…

  • Page 39
    Diagram 1: Model 240 used with the medically-approved mains adapter Mains Outlet Medical Mains Adapter Model 240 Audiometer OM024-6 Model 240 Operating Manual Page 35…
  • Page 40
    Diagram 2: Model 240 used with the medically-approved mains adapter and printer Mains Outlet Medical Mains Adapter Model 240 Audiometer Mains Outlet Printer via Printer PRINTER socket Power Supply Page 36 OM024-6 Model 240 Operating Manual…
  • Page 41
    Diagram 3: Model 240 used with the medically-approved mains adapter and PC Mains Outlet Medical Mains Adapter Model 240 Audiometer Mains Outlet Power Supply via USB socket OM024-6 Model 240 Operating Manual Page 37…

Table of Contents for AmpliVox 240:

  • OM024-6 Model 240 Operating Manual Page 35 Diagram 1: Model 240 used with the medically-approved mains adapter Mains Outlet Medical Mains Adapter Model 240 Audiometer

  • OM024-6 Model 240 Operating Manual Page 1 Model 240 Operation Manual

  • Page 12 OM024-6 Model 240 Operating Manual 3.9 Data transfer to NOAH or AudiView To transfer test results stored within the audiometer to a NOAH database the Amplivox NOAH Audilink software must be installed on to a computer. Alternatively, Amplivox AudiView allows data to be transferred to a computer and subsequently viewed, annotated & printed. This software is supplied on a CD which includes this operating manual. Refer to the installation &am

  • Contents 1 Introduction 1 1.1 Intended applications 1 1.2 Unpacking 1 1.3 Firmware version 1 1.4 Standard contents 1 1.5 Optional accessories 2 2 Important Safety Instructions 2 2.1 Precautions 2 2.2 Electromagnetic compatibility (EMC) considerations 3 2.3 Power supply options 3 2.4 Audiometer connections 4 2.5 Data transfer to a printer 5 2.6 Data transfer to a computer 5 3 Using the Audiometer 6

  • OM024-6 Model 240 Operating Manual Page 13 4.3 Headset The headset or appropriate transducer must be fitted by a qualified person to ensure a proper seal and a comfortable fit. The leads from the headset are connected to the instrument and the headset is then fitted to the patient. 4.4 Manual audiometry 4.4.1 Pre-test (1) Switch the audiometer on (2) Perform a listening check (3) Decide whether to use the manual or automatic Threshold

  • OM024-6 Model 240 Operating Manual Page 19 5.3 Physical Data Display: 2 lines of 24 characters Battery power (optional): 4×1.5V “C” cells (alkaline recommended) Mains power: 100-240Vac; 50/60Hz; 0.4A Dimensions: 270mm long x 175mm deep x 68mm high Weight: 0.75kg (approx) Safety: IEC 60601-1 (plus UL, CSA & EN deviations) EMC: IEC 60601-1-2 CE mark: To the EU Medical Device Directive 5.4 Equipmen

  • OM024-6 Model 240 Operating Manual Page 23 used once only for a single ear for a single patient. Do not reuse eartips as this will pose the risk of ear-to-ear or patient-to-patient cross infection. Further guidance is provided below:  Ensure that the black tubing protruding the foam eartip is not applied to the patient; this must be attached to the sound tube of the insert transducer  Roll the foam eartip into the smallest possible diameter 

  • Page 10 OM024-6 Model 240 Operating Manual 3.5 Threshold Retention Function This function records the thresholds for both ears at each frequency tested (air conduction and bone conduction). Thresholds may be stored manually or automatically. The operator can then review the results at the end of the test and record them on an audiogram card, print them with the optional printer (see Section 3.8), save them to the internal memory (see Section 3.6) and/or

  • OM024-6 Model 240 Operating Manual Page 11 3.6 Saving audiograms in internal memory The user may store up to 12 audiograms (AC & BC data), referenced by number, in the internal memory of the audiometer. To save the current set of audiogram thresholds (the “retained” values described in Section 3.5) press and hold the MENU key, press FREQUENCY  repeatedly until “Save Audiogram to 1” appears on screen. Use the SIGNAL keys  to select a location numbered 1-12, and then press the YES key. Rel

  • OM024-6 Model 240 Operating Manual Page 37 Diagram 3: Model 240 used with the medically-approved mains adapter and PC Mains Outlet Medical Mains Adapter Model 240 Audiometer Mains Outlet PC Power Supply PC via USB

  • Page 36 OM024-6 Model 240 Operating Manual Diagram 2: Model 240 used with the medically-approved mains adapter and printer Mains Outlet Medical Mains Adapter Model 240 Audiometer Mains Outlet Printer via Printer P

  • Page 22 OM024-6 Model 240 Operating Manual mild detergent to clean the instrument panel when required. Refer to ISO 8253-1 for additional guidance. 8.2 Transducer maintenance Before use check the transducer cables and connectors for signs of wear and/or damage. If you find any, please replace the item immediately by contacting Amplivox or your Amplivox distributor, requesting the rele

Questions, Opinions and Exploitation Impressions:

You can ask a question, express your opinion or share our experience of AmpliVox 240 device using right now.

Аудиометр Amplivox 240 – аппарат с функцией автоматического тестирования, узкополосной маскировкой, функцией сохранения порога и Talkover. Диапазон аттенюатора: до 120 дБ нПС.

Малые габариты данного устройства (270х175х68), компактный корпус с лаконичным дизайном, сравнительно небольшой вес и специальный футляр, который входит в стандартную комплектацию, позволяют ему быть одним из самых эргономичных портативных аудиометров данного производителя, которые пользуются спросом в области педиатрии, оториноларингологии, а также в общей практике.

Аппарат позволяет осуществлять аудиометрию по костной и воздушной проводимости, имеет простую систему управления, а также удобный дисплей.

Нужные данные сохраняются в памяти, переносятся на персональный компьютер или распечатываются (принтер – опциональное дополнение).

Технические характеристики аудиометра

База данных NOAH

стандарт

Функция автоматического тестирования

есть

Общие габариты

270х175х68

Принтер

Опциональное использование

АС

-10 dBHL до 120 dBHL (максимум)

BC

-10 dBHL до 70 dBHL (максимум)

Наушники

Опция

Стандарты EMC

IEC 60601-1-2

Диапазон аттенюатора

до 120 дБ нПС

Питание от батареи

опция

Оперативное запоминающее устройство

опция

Функция сохранения порога

Предназначена для автоматического или ручного сохранения пороговых значений для каждого уха и всех частот.

Точность

<1 %

Подключение

Через USB

Тесты

на костную и воздушную проводимость (с маскировкой или без маскировки).

Дисплей

С подсветкой, 2 линии 

Узкополосная маскировка

стандарт

Безопасность

IEC 60601-1

Диапазон частот

АС: 125-8kHz

BC: 250 Hz-8kHz

Используемые сигналы

Импульс, трель, сигнал одиночный.

Talkover

Специальная встроенная функция для прерывания теста и подачи голоса оператора с микрофона.

Комплектация

Стандартная комплектация аудиометра включает в себя:

  • Аудиометр
  • сетевой адаптер
  • гарнитура аудиометрическая
  • карта аудиограмм
  • футляр для транспортировки прибора
  • свидетельство о поверке
  • переключатель ответа пациента
  • гарнитура костного проводника
  • руководство по применению

Вместе с аудиометром можно также заказать дополнительные принадлежности:

  • Принтер
  • Наушник для маскировки
  • Аудиочашки (шумоподавляющие колпаки наушников)
  • Запасной кабель USB
  • Наушники-вкладыши
  • Питание от батарей
  • Интерфейсная программа NOAH (Audilink)
  • Дополнительные аудиографические карты
  • Кабель для принтера

Amplivox 240 — аудиометр диагностического типа, который предназначается для оценки слуха по дифференциальной методике. Прибор эргономичен, прост в применении, поэтому является достаточно популярным аппаратом среди потребителей. Amplivox 240 позволяет осуществлять аудиометрию по воздуху и кости. В модели есть возможность автоматического тестирования по воздуху, узкополосная маскировка, функция talk over.

Диагностический аудиометр модели 240 — это идеальный инструмент для ситуаций, когда требуется мобильность или небольшие габариты устройства. Компактный и легкий аудиометр оснащен функциями ручной и автоматической проверки воздушной и костной проводимости, а также маскировки для эффективной диагностики различных типов тугоухости. Это идеальный диагностический аудиометр, обладающий всеми необходимыми функциями для тестирования в любых условиях — дома или в дороге. Аппарат может дополнительно быть оснащен отсеком питания от батарей или же подключается к аудиометрической камере через стандартное штепсельное гнездо.

Функции, которые вам понравятся

  • Диапазон частот: 125–8000 Гц
  • Диапазон затухания: -10–100 дБ нПс
  • Воздушная (AC) и костная (BC) проводимость
  • Выбор тона: одиночный, пульсирующий, трель, постоянный
  • Узкополосная маскировка
  • Портативная и легкая конструкция
  • Дополнительная функция питания от батарей
  • Встроенная функция разговора с пациентом «Talk Forward»
  • Функция сохранения порога (во время проверки)
  • Внутренняя память
  • Включена функция простой обработки данных после тестирования

Комплектация:

  • Головные телефоны
  • Костный телефон
  • Переключатель ответа пациента
  • Бланки аудиограмм (50 штук)
  • Руководство по эксплуатации
  • Сетевой адаптер
  • Интегрированная функция общения с пациентом (Talk over)
  • Интерфейс для NOAH (для результатов в тестирования по воздуху и по кости)
  • Футляр для переноски

Дополнительные принадлежности (приобретаются отдельно)

  • Внутриушной телефон для маскировки
  • Аудиочашки (шумоподавляющие наушники)
  • Питание от батарей
  • Внутриушные телефоны
  • Внешний принтер (для распечатки аудиограммы по воздуху и кости)

Основные параметры Amplivox 240

  • Удобный в использовании
  • Компактный и портативный
  • Подключение через USB
  • Функция сохранения порога слуха
  • Внутриушные телефоны (опция)
  • 11 тестовых частот
  • Диапазон по воздуху: -10 до 120 дБ
  • Питание от батарей (опция)
  • Функция автоматического тестирования по воздуху
  • Интерфейс для NOAH

Технические характеристики

  • Вывод / ввод данных
  • Выходы:левый наушник, правый наушник, кость (левая и правая сторона), внутриушной телефон для маскировки
  • Диапазон частот (Гц):воздух:125 Гц — 8 кГц
  • кость:250 Гц — 8 кГц
  • Точность воспроизведения частоты:<1%
  • Диапазон уровня выходного сигнала (воздух):от -10 дБ нПс до 120 дБ нПс
  • Диапазон уровня выходного сигнала (кость):от -10 дБ нПс до 70 дБ нПс
  • Уровень маскировки:90 дБ нПс макс (250 Гц – 4 кГц)
  • Выходной сигнал — точность:в пределах 3 дБ
  • Выходной сигнал — шаг:5 дБ
  • Преобразователь выходного сигнала:стандартные головные телефоны и/или внутриушные телефоны(опция)
  • Тип сигнала:одиночный, частотно-модулированный или импульсный
  • Тестирование по воздуху:ручной или автоматический режим
  • Маскировка:узкополосный шум
  • Связь:встроенное устройство для переговоров
  • Интерфейс для NOAH:передача порогов слуха по воздуху и кости.

Технические характеристики

Страна производитель:

Англия

Гарантийный срок:

12 месяцев

Частотный диапазон (Гц):

125 -8000

При покупке данной продукции вы получите:

  1.   Гарантийный талон
  2.   Паспорт аппарата
  3.   Инструкцию по эксплуатации
  4.   Сертификат качества
  5.   Регистрационное удостоверение
  6.   Товарную накладную
  7.   Кассовый чек

Amplivox 240 является диагностическим аудиометром эконом-класса, который идеально подходит для поликлиник, а также мобильного использования.


Задать вопрос

Аудиометр Amplivox 240

Аудиометр Amplivox 240

Amplivox 240 является диагностическим аудиометром эконом-класса, который идеально подходит для поликлиник, а также мобильного использования.


Задать вопрос

Skip to navigation
Skip to content

Поставка нового и восстановленного медицинского оборудования

Распродажа!

274 975 

-15%

Тип прибора: Диагностический
Тесты: костная и воздушная проводимость
Маскировка: есть
База данных NOAH: есть
Функция автоматического тестирования: есть

Доставка

По всей территории России проверенными ТК

Поддержка

Послепродажная поддержка и обслуживание клиентов

Производители

Лучшие мировые бренды медицинского оборудования

Документы

На всё оборудование есть необходимая документация

  • Главная
  • Продукция
  • ЛОР оборудование
  • Аудиометры
  • Аудиометр диагностический Amplivox 240, Англия



Цена:
323 500
руб.


В наличии
2 шт.


  • Москва: 2


Страна-производитель: Англия

Amplivox 240 — аудиометр диагностического типа, который предназначается для оценки слуха по дифференциальной методике. Прибор эргономичен, прост в применении, поэтому является достаточно популярным аппаратом среди потребителей.

Эргономика

Все элементы для управления и работы с прибором расположены максимально удобно для пользователя. Результаты, которые получаются в итоге исследования, отражаются на дисплее моментально, что позволяет оценить состояние пациента оперативно.

Простое использование

Аппарат:

  • имеет функцию сохранения порогов слуха, что минимизирует время тестирования пациента;
  • позволяет максимально концентрировать внимание пациента;
  • имеет бесщелчковые аттенюаторы;
  • итоги исследовательского мероприятия сохраняются в памяти аппарата при желании пользователя;
  • результаты можно вывести на печать или передать на ПК;
  • компактный, портативный, современный;
  • можно использовать и в условиях стационара, и в полевых зонах;
  • поставка прибора осуществляется с особым кейсом.

Комплектация

  • Головные телефоны
  • Костный телефон
  • Кнопка ответа пациента
  • Бланки аудиограмм (50 штук)
  • Сетевой адаптер
  • Кейс для переноски
  • Программное обеспечение Amplisuite на русском языке
  • Руководство по эксплуатации на русском языке

Технические характеристики

Выходы Левый наушник, правый наушник, кость (левая и правая сторона), внутриушной телефон для маскирования
Диапазон частот (Гц) Воздух: 125 Гц — 8 кГц
Кость: 250 Гц — 8 кГц
Точность воспроизведения частоты <1%
Диапазон уровня выходного сигнала (воздух) от -10 дБ нПс до 120 дБ нПс (максимум)
Диапазон уровня выходного сигнала (кость) от -10 дБ нПс до 70 дБ нПс (максимум)
Уровень маскировки 90 дБ нПс макс (250 Гц – 4 кГц)
Выходной сигнал — точность в пределах 3 дБ
Выходной сигнал — шаг 5 дБ
Тип сигнала одиночный, частотно-модулированный или импульсный
Маскировка узкополосная
Связь Встроенный микрофон
Дисплей 2 строки по 24 символа в каждой
Питание от сети 100 – 240 В перем. тока; 50/60 Гц;
Габариты прибора длина 270 мм x ширина 175 мм x высота 68 мм, вес — 735г

Понравилась статья? Поделить с друзьями:

А вот и еще наши интересные статьи:

  • Мануал по двигателю 4hk1
  • Lavr ml100 petrol инструкция по применению
  • Нео трависил леденцы инструкция по применению детям
  • Принтер кэнон i sensys mf3010 инструкция по применению
  • Инструкция для лица ответственного за организацию эксплуатации лифтов

  • 0 0 голоса
    Рейтинг статьи
    Подписаться
    Уведомить о
    guest

    0 комментариев
    Старые
    Новые Популярные
    Межтекстовые Отзывы
    Посмотреть все комментарии